G-ARMOR Stent® Portfolio

Product Quick Finder

Choose a category or subcategory

The G-ARMOR Stent®, part of the CP Stent® Family, is the latest large diameter, balloon expandable stent technology.

  • Composed of platinum-iridium wire arranged in a unique, laser welded 8 or 10 “zig” pattern and over brazed with 24K gold.
  • Available in a Bare and an Expandable ePTFE Covered version in both Pre-Mounted (on a BIB® Catheter) and Unmounted configurations.
  • The G-ARMOR Stent® is compatible with introducers ranging from 13 - 18F.
  • 12.0mm – 30.0mm stent diameters
  • 4.6cm – 6.3cm stent lengths

Indications for Use

The G-ARMOR Stent® is indicated for use in the treatment of native and/or recurrent coarctation of the aorta involving a compliant aortic isthmus or first segment of the descending aorta where there is adequate size and patency of at least one femoral artery and balloon angioplasty is contraindicated or predicted to be ineffective.

The G-ARMOR Covered Stent™ is indicated for use in the treatment of native and/or recurrent coarctation of the aorta involving the aortic isthmus or first segment of the descending aorta where there is adequate size and patency of at least one femoral artery associated with one or more of the following: acute or chronic wall injury; nearly atretic descending aorta of 3 mm or less in diameter; a non-compliant stenotic aortic segment found on pre-stent balloon dilation; a genetic or congenital syndrome associated with aortic wall weakening or ascending aortic aneurysm. The G-ARMOR Covered Stent™ is indicated for use in the treatment of right ventricle to pulmonary artery (right ventricular outflow tract) conduit disruptions that are identified during conduit pre-dilatation procedures performed in preparation for transcatheter pulmonary valve replacement.

Caution: Federal (USA) Law restricts this device to sale by or on the order of a physician. Contraindications: Clinical or biological signs of infection. Active endocarditis. Pregnancy. Contraindications (CoA only): Patients too small to allow safe delivery of the stent without compromise to the systemic artery used for delivery. Unfavorable aortic anatomy that does not dilate with high pressure balloon angioplasty. Curved vasculature. Occlusion or obstruction of systemic artery precluding delivery or the stent. Known allergy to aspirin, other antiplatelet agents, or heparin. Contraindications (RVOT only): Patients too small to allow safe delivery of the stent without injury to a systemic vein or to the right side of the heart. Warnings / Precautions: Radiofrequency heating during MRI scans on overlapped, 10 zig G-ARMOR stents has not been evaluated. Excessive force while crimping may weaken welds of the stent. Crimping the 8 zig stent on a balloon catheter smaller than 12mm, and the 10 zig on a balloon catheter smaller than 26mm, may cause damage to the stent. The stent is rigid and may make negotiation through vessels difficult. Warnings / Precautions (CoA only): Coarctation of the aorta involving the aortic isthmus or first segment of the descending aorta should be confirmed by diagnostic imaging. The NuMED G-ARMOR Stent has not been evaluated in patients weighing less than 20kg. As with any type of implant, infection secondary to contamination of the stent may lead to aortitis, or abscess. Over-stretching of the artery may result in rupture or aneurysm formation. Warnings / Precautions (G-ARMOR Covered Stent only): Excessive handling and manipulation of the covering while crimping the stent may cause the covering to tear off of the stent. Crimping the device in the opposite direction of the folds in the covering may cause the covering to catch while inserting into the hemostasis tool and introducer. This could cause the covering to tear off the stent. Pulling the Covered stent back through the introducer and/or hemostasis valve may cause the covering to catch and tear off of the stent. Warnings / Precautions (RVOT only): During the Premarket Approval study the Medtronic Melody valve was used for valve restoration. The safety and effectiveness of the Covered G-ARMOR Stent for pre-stenting of the right ventricular outflow tract (RVOT) landing zone (i.e. prophylaxis or prevention of either RVOT conduit rupture or TPVR fracture; use as a primary RVOT conduit) in preparation of a transcatheter pulmonary valve replacement (TPVR) has not been evaluated. As with any type of implant, infection secondary to contamination of the stent might lead to endocarditis, or abscess formation. The Covered Stent can migrate from the site of implant potentially causing obstruction to pulmonary artery flow. Over-stretching of the RVOT may result in rupture or aneurysm of the RV-PA conduit or the native pulmonary artery. The inflated diameter of the stent should at least equal the diameter of the intended implant site. and/or hemostasis valve may cause the covering to catch and tear off of the stent. 

Articles

Description Article Code
G ARMOR BARE STENT G10Z46 614715
G ARMOR BARE STENT G10Z52 614716
G ARMOR BARE STENT G10Z57 614717
G ARMOR BARE STENT G10Z63 614718
G ARMOR BARE STENT G8Z46 614719
G ARMOR BARE STENT G8Z52 614720
G ARMOR BARE STENT G8Z57 614721
G ARMOR BARE STENT G8Z63 614722
G ARMOR GCM001 4.6CM 12MM 5CM 14F 614723
G ARMOR GCM002 4.6CM 14MM 5CM 14F 614724
G ARMOR GCM003 4.6CM 15MM 5CM 14F 614725
G ARMOR GCM004 4.6CM 16MM 5CM 14F 614726
G ARMOR GCM005 4.6CM 18MM 5CM 14F 614727
G ARMOR GCM006 4.6CM 20MM 5CM 16F 614728
G ARMOR GCM007 4.6CM 22MM 5CM 16F 614729
G ARMOR GCM008 4.6CM 24MM 5CM 16F 614730
G ARMOR GCM009 5.2CM 12MM 5.5CM 14F 614731
G ARMOR GCM010 5.2CM 14MM 5.5CM 14F 614732
G ARMOR GCM011 5.2CM 15MM 5.5CM 14F 614733
G ARMOR GCM012 5.2CM 16MM 5.5CM 14F 614734
G ARMOR GCM013 5.2CM 18MM 5.5CM 14F 614735
G ARMOR GCM014 5.2CM 20MM 5.5CM 16F 614736
G ARMOR GCM015 5.2CM 22MM 5.5CM 16F 614737
G ARMOR GCM016 5.2CM 24MM 5.5CM 16F 614738
G ARMOR GCM017 5.7CM 12MM 6CM 14F 614739
G ARMOR GCM018 5.7CM 14MM 6CM 14F 614740
G ARMOR GCM019 5.7CM 15MM 6CM 14F 614741
G ARMOR GCM020 5.7CM 16MM 6CM 14F 614742
G ARMOR GCM021 5.7CM 18MM 6CM 14F 614743
G ARMOR GCM022 5.7CM 20MM 6CM 16F 614744
G ARMOR GCM023 5.7CM 22MM 6CM 16F 614745
G ARMOR GCM024 5.7CM 24MM 6CM 16F 614746
G ARMOR GCM025 6.3CM 12MM 6.5CM 14F 614747
G ARMOR GCM026 6.3CM 14MM 6.5CM 14F 614748
G ARMOR GCM027 6.3CM 15MM 6.5CM 14F 614749
G ARMOR GCM028 6.3CM 16MM 6.5CM 14F 614750
G ARMOR GCM029 6.3CM 18MM 6.5CM 14F 614751
G ARMOR GCM030 6.3CM 20MM 6.5CM 16F 614752
G ARMOR GCM031 6.3CM 22MM 6.5CM 16F 614753
G ARMOR GCM032 6.3CM 24MM 6.5CM 16F 614754
G ARMOR GCM033 4.6CM 26MM 5CM 16F 614755
G ARMOR GCM034 4.6CM 28MM 5CM 18F 614756
G ARMOR GCM035 4.6CM 30MM 5CM 18F 614757
G ARMOR GCM036 5.2CM 26MM 5.5CM 16F 614758
G ARMOR GCM037 5.2CM 28MM 5.5CM 18F 614759
G ARMOR GCM038 5.2CM 30MM 5.5CM 18F 614760
G ARMOR GCM039 5.7CM 26MM 6CM 16F 614761
G ARMOR GCM040 5.7CM 28MM 6CM 18F 614762
G ARMOR GCM041 5.7CM 30MM 6CM 18F 614763
G ARMOR GCM042 6.3CM 26MM 6.5CM 16F 614764
G ARMOR GCM043 6.3CM 28MM 6.5CM 18F 614765
G ARMOR GCM044 6.3CM 30MM 6.5CM 18F 614766
G ARMOR GM001 4.6CM 12MM 5CM 13F 614767
G ARMOR GM002 4.6CM 14MM 5CM 13F 614768
G ARMOR GM003 4.6CM 15MM 5CM 13F 614769
G ARMOR GM004 4.6CM 16MM 5CM 14F 614770
G ARMOR GM005 4.6CM 18MM 5CM 14F 614771
G ARMOR GM006 4.6CM 20MM 5CM 14F 614772
G ARMOR GM007 4.6CM 22MM 5CM 14F 614773
G ARMOR GM008 4.6CM 24MM 5CM 14F 614774
G ARMOR GM009 5.2CM 12MM 5.5CM 13F 614775
G ARMOR GM010 5.2CM 14MM 5.5CM 13F 614776
G ARMOR GM011 5.2CM 15MM 5.5CM 13F 614777
G ARMOR GM012 5.2CM 16MM 5.5CM 14F 614778
G ARMOR GM013 5.2CM 18MM 5.5CM 14F 614779
G ARMOR GM014 5.2CM 20MM 5.5CM 14F 614780
G ARMOR GM015 5.2CM 22MM 5.5CM 14F 614781
G ARMOR GM016 5.2CM 24MM 5.5CM 14F 614782
G ARMOR GM017 5.7CM 12MM 6CM 13F 614783
G ARMOR GM018 5.7CM 14MM 6CM 13F 614784
G ARMOR GM019 5.7CM 15MM 6CM 13F 614785
G ARMOR GM020 5.7CM 16MM 6CM 14F 614786
G ARMOR GM021 5.7CM 18MM 6CM 14F 614787
G ARMOR GM022 5.7CM 20MM 6CM 14F 614788
G ARMOR GM023 5.7CM 22MM 6CM 14F 614789
G ARMOR GM024 5.7CM 24MM 6CM 14F 614790
G ARMOR GM025 6.3CM 12MM 6.5CM 13F 614791
G ARMOR GM026 6.3CM 14MM 6.5CM 13F 614792
G ARMOR GM027 6.3CM 15MM 6.5CM 13F 614793
G ARMOR GM028 6.3CM 16MM 6.5CM 14F 614794
G ARMOR GM029 6.3CM 18MM 6.5CM 14F 614795
G ARMOR GM030 6.3CM 20MM 6.5CM 14F 614796
G ARMOR GM031 6.3CM 22MM 6.5CM 14F 614797
G ARMOR GM032 6.3CM 24MM 6.5CM 14F 614798
G ARMOR GM033 4.6CM 26MM 5CM 16F 614799
G ARMOR GM034 4.6CM 28MM 5CM 16F 614800
G ARMOR GM035 4.6CM 30MM 5CM 16F 614801
G ARMOR GM036 5.2CM 26MM 5.5CM 16F 614802
G ARMOR GM037 5.2CM 28MM 5.5CM 16F 614803
G ARMOR GM038 5.2CM 30MM 5.5CM 16F 614804
G ARMOR GM039 5.7CM 26MM 6CM 16F 614805
G ARMOR GM040 5.7CM 28MM 6CM 16F 614806
G ARMOR GM041 5.7CM 30MM 6CM 16F 614807
G ARMOR GM042 6.3CM 26MM 6.5CM 16F 614808
G ARMOR GM043 6.3CM 28MM 6.5CM 16F 614809
G ARMOR GM044 6.3CM 30MM 6.5CM 16F 614810
G ARMOR COVERED GC10Z46 614811
G ARMOR COVERED GC10Z52 614812
G ARMOR COVERED GC10Z57 614813
G ARMOR COVERED GC10Z63 614814
G ARMOR COVERED GC8Z46 614815
G ARMOR COVERED GC8Z52 614816
G ARMOR COVERED GC8Z57 614817
G ARMOR COVERED GC8Z63 614818